Model Number CI-1400-01 |
Device Problems
Mechanical Problem (1384); Fitting Problem (2183)
|
Patient Problems
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however, the issue did not resolve.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed tool damage to the top and bottom covers, as well as a severed electrode.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical and mechanical tests performed.The internal visual inspection noted silver migration between an electrical component.Based on an assessment of the helium leak test data and internal visual inspection, it is determined that this device was non-hermetic, and that the root cause of the excessive moisture was a leak through the feedthru seals, even though the rga results were within specifications.A corrective action was implemented.This version of the 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|