MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA ONE 3.1
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2021
Event Type  malfunction  
Manufacturer Narrative
The metal plate underneath the support arm is used to index the lock system.This one fits in the oblong slots which allows the locking of the telescopic arm.The system was compliant at the time of its release and the assembled screws are coated with threadlocker and are tightened to 2.2n.M.It is unlikely that these screws could have fallen by themselves, someone certainly tried to disassemble the metal plate by unscrewing these 5 screws.The new screws were replaced on (b)(6) 2021 and no surgery was performed before this replacement.Unique identifier (udi) #: (b)(4).
 
Event Description
The clinical representative (cr) was present for a preventive maintenance.During the maintenance, cr noticed that the metal plate underneath the support arm had lost 5 screws, and was not completely seated on the support arm when the arm is not locked in place.When the arm is locked in place, the support arm is still rigid and does not move.The robot passed applicative and accuracy tests and is still in conformance for use during procedures.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key11715391
MDR Text Key247280889
Report Number3009185973-2021-00112
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200511
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA ONE 3.1
Device Catalogue NumberROSAS00203
Device Lot Number3.1.4.1650
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/28/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-