H4: the lot was manufactured from march 16, 2020 - march 18, 2020.H10: the actual devices were not available; however, two (2) photographs of the samples were provided for evaluation.The photographs were visually inspected and the white flow restrictor was observed cracked which suggested the cause of the leak.The photographs also showed a non-baxter connector attached to the white flow restrictor.When the non-baxter connector does not properly fit the flow restrictor, it can cause a crack on the flow restrictor which would result in a leak.The reported condition was verified.The cause of the condition was not determined, however the most probable cause of the cracked luer which resulted in a leak is due to the non-baxter connector.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|