Model Number 3100A
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2021
Event Type  malfunction  
Manufacturer Narrative
A vyaire field service representative (fsr) went onsite and evaluated the ventilator.Opened unit up and checked for any burn damage and did not see any charred or burned components.Tested thermistor by doing final test on the 6k pm and the thermistor was functional.Completed functional check out and unit is working.Unit requires new castor.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported to vyaire medical that the 3100a ventilator has an overheat issue while on patient.The customer confirmed that there was no patient harm associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, IL 60045
8333273284
MDR Report Key11721218
MDR Text Key247801627
Report Number2021710-2021-13768
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20040429
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-