MDT SOFAMOR DANEK PUERTO RICO MFG VENTURE ANTERIOR CERVICAL PLATE SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL
|
Back to Search Results |
|
Model Number 9790313 |
Device Problems
Degraded (1153); Device Slipped (1584)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from healthcare professional via field representative regarding patient with symptoms of cervical spinal stenosis involved in extended decompression of cervical spinal canal, bone graft fusion and internal fixation procedure.It was reported that at intra-op, screw worn out and screw's tip slipped.Event was associated with the patient.Product was utilized correctly according to the directions given in the ifu/labeling.There were no patient symptoms or complications reported as a result of this event.There was no delay in overall procedure time.Patient is alive and no injury.
|
|
Manufacturer Narrative
|
H3: device evaluated summary- visual and optical inspection confirmed the female hex of the screw has been stripped.The edges have been deformed and rounded off.This type of damage is consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|
|
|