SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
|
Back to Search Results |
|
Model Number 71118200 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/05/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during inspection the trigen reamer set had a flexible shaft w/cir connector that was cracked (case: (b)(4)), and the reamers 9 mm, 10.5 mm, 11 mm, 13 mm, and 14 mm were dull (case: (b)(4)).No case reported; therefore, there was no patient involvement.
|
|
Manufacturer Narrative
|
The device, used in treatment, was returned for evaluation.A visual inspection was performed and confirmed the flexible shaft with connector is broken on the top of the device.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|