MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATUN REV LPS INS BASE TRL ADPT; TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATUN REV LPS INS BASE TRL ADPT; TIBIAL TRIALS Back to Search Results
Model Number 2517-60-002
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that upon removal of insert trial it was noticed that the spring on the metal trial broke.All pieces were retrieved.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned instrument confirmed the event of device breakage.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUN REV LPS INS BASE TRL ADPT
Type of Device
TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11743814
MDR Text Key248832416
Report Number1818910-2021-09128
Device Sequence Number1
Product Code NJL
UDI-Device Identifier10603295480174
UDI-Public10603295480174
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2517-60-002
Device Catalogue Number251760002
Device Lot NumberMI84231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Date Manufacturer Received06/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-