Model Number CI-1600-04 |
Device Problems
Mechanical Problem (1384); Inaudible or Unclear Audible Prompt/Feedback (2283); Impedance Problem (2950)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 04/22/2021 |
Event Type
malfunction
|
Event Description
|
A review of the recipient¿s test data indicated impedance issues.The recipient is presenting with sound quality issues.Programming adjustments were made, however, the issues did not resolve.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted.The recipient was re-implanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The device passed the external visual inspection.The photographic imaging inspection revealed some electrode were broken within the electrode pocket.System lock was verified.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|