Model Number EV08182CD |
Device Problem
Component Missing (2306)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 03/2022).
|
|
Event Description
|
It was reported that prior to a stent placement, the stent allegedly missing from the package.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 03/2022).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that prior to a stent placement, the stent allegedly missing from the package.There was no patient contact.
|
|
Event Description
|
It was reported that prior to a stent placement, the stent allegedly missing from the package.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one valeo pta dilatation catheter has been received for the evaluation.On the visual evaluation, the sample appeared clean.Under the microscopic observation, no stent was noted and the crimping marks were observed on the balloon.Furthermore, no kinks were noted to the catheter and no anomalies were observed on the luers or y-body.No functional testing was performed due to the reported nature of the failure.Therefore the investigation for the reported stent missing was inconclusive as the stent noted to be missed from the returned device.However the crimping mark was noted on the balloon.A definitive root cause for the stent missing could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiry date: 03/2022), h11: h6 (method), h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
|