Model Number 407645
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Connection Problem (2900); Impedance Problem (2950); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the right atrial (ra) lead exhibited high thresholds and varying, spiking and high impedance.It also noted that lead integrity issue is suspected.The ra lead remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient's right atrial (ra) lead exhibited high and spiked pacing impedance, as well as high thresholds.Upon arm manipulation, the lead also exhibited noise consistent with a connection issue.The patient elected not to have surgery due to imminent relocation, therefore the lead alert was turned off and both the implantable cardioverter defibrillator (icd) and ra lead remain in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS MRI SURESCAN
Type of Device
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key11775700
MDR Text Key249011182
Report Number2649622-2021-08957
Device Sequence Number1
Product Code DTB
UDI-Device Identifier00643169708181
UDI-Public00643169708181
Combination Product (y/n)N
PMA/PMN Number
P930039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/04/2022
Device Model Number407645
Device Catalogue Number407645
Was Device Available for Evaluation? No
Date Manufacturer Received04/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
6935M55 LEAD, DDPB3D4 ICD; 6935M55 LEAD, DDPB3D4 ICD
Patient Age75 YR