Model Number N/A
Device Problems Failure to Calibrate (2440); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor speed was low and the unit was out of calibration at the zero setting.The motor was replaced and the unit was calibrated which resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that during preventative maintenance the device had a defective motor that needed replaced and the device needed calibrated.No further information provided.Investigation found the motor speed was low.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11776270
MDR Text Key249295852
Report Number0001526350-2021-00524
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375901
UDI-Public(01)00889024375901(11)190121(10)64277076
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number64277076
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2021
Was the Report Sent to FDA? No
Date Manufacturer Received04/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-