Model Number A3059
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2021
Event Type  malfunction  
Manufacturer Narrative
Device history record: the dhr shows no abnormalities related to the reported failure.Mayfield composite series skull clamp (a3059) was returned for evaluation.The reported complaint was confirmed via inspection of the unit.Unit received showing signs of wear, the lock had an up and down movement and the teeth were grinding when rotated.General maintenance and cleaning will be performed to resolve observed issues.
 
Event Description
A facility reported that the locking knob on the mayfield skull clamp (a3059) became stuck and was impossible to unlock to adjust the clamp on the patient.There was a 15 minute delay in surgery with no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key11790051
MDR Text Key250184052
Report Number3004608878-2021-00325
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253792
UDI-Public10381780253792
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA3059
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2021
Date Manufacturer Received04/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-