Model Number 71992-01 |
Device Problems
Premature End-of-Life Indicator (1480); Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Diaphoresis (2452); Confusion/ Disorientation (2553)
|
Event Date 05/02/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kit were reviewed and the dhrs showed the libre sensor and libre sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
Caller reported the customer's freestyle libre 2 sensor expired prematurely and customer was unable to obtain readings.As a result, the customer experienced symptoms described as confusion, pallor, and sweating and was unable to self-treat.Customer's wife treated the customer with a glucagon injection and no further treatment was reported.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug is properly seated and no physical damage was observed on sensor patch.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Sim vivo testing (simulation of the electrical signal produced by the sensor tail) was performed and all results were within specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
Caller reported the customer's freestyle libre 2 sensor expired prematurely and customer was unable to obtain readings.As a result, the customer experienced symptoms described as confusion, pallor, and sweating and was unable to self-treat.Customer's wife treated the customer with a glucagon injection and no further treatment was reported.There was no report of death or permanent injury associated with this event.
|
|