SMITH & NEPHEW, INC. GII INTRA TIB ALIGNMENT; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71440200
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2021
Event Type  malfunction  
Event Description
It was reported that, during a tka, a gii intra tib alignment had difficult to pass through.The procedure was able to be completed with the same device.Surgery was not delayed.Patient was not harmed.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection was performed and confirmed the intra tib alignment has a damaged bar on the reamer holder sub-assembly of the device.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII INTRA TIB ALIGNMENT
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11797850
MDR Text Key249717742
Report Number1020279-2021-03912
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010214423
UDI-Public03596010214423
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440200
Device Catalogue Number71440200
Device Lot Number80305539
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-