• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES RI ROBOTIC DRILL ATTACHMENT; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES RI ROBOTIC DRILL ATTACHMENT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10015
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2021
Event Type  malfunction  
Event Description
It was reported that, after a cori tka, they noticed that the white part (scraper at the end of the attachment, inside) was separated from the robotic drill attachment.Since the incident occurred after the procedure, there was no patient involvement.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the cori drill attachment p/n rob10015 (b)(6), intended for use in treatment was returned.The reported problem was confirmed.The wiper is separated from the drill attachment.No epoxy is visible around the crown for adhesion.Although the reported problem was visually confirmed a functional evaluation was performed.A drill test was run and the passed.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event was lack of epoxy applied during manufacturing.Based on the investigation, no further containment or corrective action is recommended or required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RI ROBOTIC DRILL ATTACHMENT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11803051
MDR Text Key249955127
Report Number3010266064-2021-00348
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757345
UDI-Public00885556757345
Combination Product (y/n)N
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10015
Device Catalogue NumberROB10015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2021
Date Manufacturer Received06/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ROB10024, SERIAL: (B)(6)
-
-