Product analysis: analysis was able to confirm the customer comment that the atrial and/or ventricular patient cable connectors on the external pulse generator (epg) were broken, cracked or defective as the output connector assembly was broken.Analysis determined that epg failed the incoming functional test "ventricular pace pulse noise" as the main printed circuit board (pcb) was out of specification electrical and the display wire was pinched with wire insulation exposed.It was noted that the hanger assembly and the upper case was broken.All found defective parts were replaced and all other identified issues were resolved.The epg then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|
This complaint was confirmed based on the actual device.Benchtop analysis was performed the main board was assembled into a golden unit.The ventricular pace pulse noise was measured and was found to be within the requirements of the ventricular pace pulse noise test specification.There was no defect found.The instrument pc board failure was not confirmed.The most likely root cause was traced to a component failure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|