VENTED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290V
Device Problems Crack (1135); Leak/Splash (1354); Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We are currently in the process of finalising our investigation.We will provide a follow up report upon completion of investigation.
 
Event Description
A distributor in (b)(6) reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that a mr290v vented autofeed humidification chamber was leaking from the chamber base prior to patient use.There was no patient involvement.
 
Manufacturer Narrative
Ps367595 correction: section b5 and h6 were corrected.Method: the complaint mr290v vented autofeed humidification chamber was returned to fisher & paykel healthcare (f&p) in new zealand where it was visually inspected and analysed.Results: : visual inspection of the returned mr290v chamber revealed a vertical crack line on the bottom of the chamber dome.Conclusion: we are unable to determine the root cause of the chamber crack, however our investigation indicates that the crack is most likely due to mechanical stress.The stress source was unable to be identified.The mr290 chambers are designed and tested to conform to iso 5367 breathing tubes intended for use with anaesthetic apparatus and ventilators.Every mr290 chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber dome due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions, could lead to a loss of ventilation pressure".
 
Event Description
A distributor in japan reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that a mr290v vented autofeed humidification chamber was leaking water from the chamber base during use.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
MDR Report Key11810466
MDR Text Key252739040
Report Number9611451-2021-00564
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 04/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2021
Date Manufacturer Received08/18/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
F&P MR850 RESPIRATORY HUMIDIFIER; F&P MR850 RESPIRATORY HUMIDIFIER; RT228 INFANT BREATHING CIRCUIT; SLE6000 VENTILATOR; SLE6000 VENTILATOR
-
-