MAUDE Adverse Event Report: IMPRESS®.BRAIDED CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPRESS®.BRAIDED CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 53035KA2
Device Problems Material Rupture (1546); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2021
Event Type  malfunction  
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that during a peripheral vascular procedure, the catheter tip detached outside the patient.The physician had successfully acquired retrograde arterial access, and the catheter was introduced over a stiff guidewire, the lesion was crossed, and contrast was injected to collect selective angiograms.The catheter was removed from the patient and placed back on the sterile field.At a later point within the procedure, the clinician decided to use the 5f ka2 catheter again.When an attempt was made to reload the catheter over the back end of the guidewire, the tip detached and was dangling from the catheter shaft.A new catheter was opened and used to complete the procedure.No patient injury to report.
 
Manufacturer Narrative
The suspect device has been returned for evaluation.The complaint is confirmed.The root cause is attributed to the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS®.BRAIDED CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
MDR Report Key11813233
MDR Text Key250238231
Report Number3010665433-2021-00027
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00884450032565
UDI-Public884450032565
Combination Product (y/n)N
PMA/PMN Number
K053171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/18/2024
Device Catalogue Number53035KA2
Device Lot NumberE2065224
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2021
Date Manufacturer Received07/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LAURATE STIFF GUIDEWIRE.; LAURATE STIFF GUIDEWIRE
-
-