The evaluation of the failed scope has yielded no cause due to product manufacture or design.The evaluation also found the angle cover was melted; the working channel has a leak; the shaft is bent; dents in the tip; and the handle housing was found soiled.The device shows signs of maintenance and care issues.Further testing provided the electronics within the shaft is incapable of getting hot enough to have melted the angle cover material.Note: a new report was needed instead of a supplement due to the need to correct the manufacturer's number, in the mdr report number, and also the manufacturer's name.
|
During an endoscopic 3rd ventriculostomy, while removing the scope at the end of procedure, there seemed to be resistance.Tissue was found to be adhered to the side of scope; cautery was used to remove scope.The procedure was completed with no other surgical intervention required and, per dr.(b)(6), the patient does not have any neurologic consequences.
|