• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems Intermittent Capture (1080); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/15/2021
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was unable to confirm the customer comments: the device passed the bench tests but the device failed ventricular pace pulse noise at incoming functional testing.It was reported that the display wires were pinched with the wire exposed.It was noted that the hanger assembly was cracked and the lower case was damaged next to the battery drawer.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that there was an intermittent loss of capture with the external pulse generator (epg) and it was not capturing the patient's heart beat consistently.It was unknown if the loss of capture was on the right ventricular (rv) or right atrial (ra) lead.It was noted that the epg unit was changed out and the issue was resolved.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Analysis of the main board found no visual anomalies.The device failed ventricular pace pulse noise at the incoming functional testing.At benchtop analysis the ventricular pace pulse noise measured within requirements.No defect found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key11819428
MDR Text Key250444533
Report Number2182208-2021-01937
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169508767
UDI-Public00643169508767
Combination Product (y/n)N
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Date Manufacturer Received05/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age51 YR
Patient Weight115
-
-