Model Number 560BCS1 |
Device Problem
Insufficient Flow or Under Infusion (2182)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of this bio-console instrument during an extra corporeal membrane oxygenation (ecmo) case using the instrument and motor after 90 minutes of use, the patient was assembled and ecmo support was started with the new equipment, and a low flow alarm presented at the half hour.The customer stated that there was no presence of cavitation, or obstruction in the flow and the flowmeter was properly placed.The customer did not observe changes in the rpm, so they changed to the manual motor (hand crank), and then the external motor was changed and the support was reset with the patient at 65%, during the event.No hemodynamic changes were observed.Use of instrument was unspecified and there was no adverse patient effect associated with this event.Additional information was received on 15 april 2021 stating that the hand crank was used for less than 30 minutes.
|
|
Manufacturer Narrative
|
Product analysis: the reported low flow alarm was verified during field service.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis: the bio-console instrument underwent further analysis and the reported low flow alarm was not verified during service.The service technician found no issue with the instrument and they stated that the reported issue was not associated with the instrument.Preventive maintenance was performed as per specification.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|