This event has been recorded by zimmer biomet under (b)(4).Functional testing of the returned product determined the device would power on and function in low mode, but not high mode when connected to the returned battery pack.Visual examination of the returned product identified the device was used; fluid in the tubing and blood on the handgun.One of the batteries had leaked and crystallized in the terminal.Once the terminal was cleared of the debris and the faulty battery was replaced, the device powered on and functioned properly in both modes.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|