MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HIP KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HIP KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problems Expulsion (2933); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/24/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Functional testing of the returned product determined the device would power on and function in low mode, but not high mode when connected to the returned battery pack.Visual examination of the returned product identified the device was used; fluid in the tubing and blood on the handgun.One of the batteries had leaked and crystallized in the terminal.Once the terminal was cleared of the debris and the faulty battery was replaced, the device powered on and functioned properly in both modes.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that the work time of pulsavac was short and water pressure was low when the surgeon turned on the switch.Investigation found the battery had leaked.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11825288
MDR Text Key250871187
Report Number0001526350-2021-00567
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00889024375192
UDI-Public(01)00889024375192(17)221231(10)6463982
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberN/A
Device Catalogue Number00515048201
Device Lot Number6463982
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2021
Was the Report Sent to FDA? No
Date Manufacturer Received05/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-