• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR - REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR - REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the cart would not run on and the power inlet module was bad and has melted, both fuses were also blown and the power cord was damaged.The power cord and inlet module were replaced and resolved the reported issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item.However, as the lot number is unknown, an additional review could not be performed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported the cart is not powering on during cleaning.The power inlet module melted and the head of the power cord that inserts into the power inlet module was charred from the event.There were no frayed or exposed wires.No adverse events were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR - REFURBISHED
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11838047
MDR Text Key257541069
Report Number0001954182-2021-00025
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024466029
UDI-Public(01)00889024466029
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUE514010200
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-