This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the cart would not run on and the power inlet module was bad and has melted, both fuses were also blown and the power cord was damaged.The power cord and inlet module were replaced and resolved the reported issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item.However, as the lot number is unknown, an additional review could not be performed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|