• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETERS Back to Search Results
Model Number CRUS6106
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2021
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer.The investigation of the reported event is currently underway.(expiration date: 10/2022).
 
Event Description
It was reported that during a recanalization procedure, the tip of the catheter allegedly detached.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one 6s crosser cto recanalization catheter was returned for evaluation.It was noted the distal tip was detached from the catheter and was not returned with sample.The distal marker band was not attached, outer catheter is also noted to be stretched.Distal tip detachment measures 105.5 cm from break to strain relief.Based on the findings, the investigation is confirmed for the reported detachment issue as the it was noted hat the distal tip was completely detached form the catheter.A definitive root cause could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiration date: 10/2022),g3, h6(method).H11: b5, h1, h6(patient, result and conclusion).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that during a recanalization procedure, the tip of the catheter allegedly detached.It was further reported that the tip was left in the calcified lesion.The current status of the patient was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETERS
Type of Device
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11839990
MDR Text Key251187612
Report Number2020394-2021-01055
Device Sequence Number1
Product Code PDU
UDI-Device Identifier10801741125642
UDI-Public(01)10801741125642
Combination Product (y/n)N
PMA/PMN Number
K161208
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCRUS6106
Device Catalogue NumberCRUS6106
Device Lot NumberGFEY2153
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2021
Date Manufacturer Received06/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age74 YR
-
-