U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY MARCON-HABER VARICES INJECTOR; FBK, ENDOSCOPIC INJECTION NEEDLE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY MARCON-HABER VARICES INJECTOR; FBK, ENDOSCOPIC INJECTION NEEDLE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number G21655
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter section: occupation - unknown.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all marcon-haber varices injector are subjected to a visual inspection and functional testing to ensure device integrity.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic procedure, the physician used a cook marcon-haber varices injector.The needle passed through the endoscope but the tip did not come out.Upon collecting and removing from the device, it was identified that the needle was broken.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARCON-HABER VARICES INJECTOR
Type of Device
FBK, ENDOSCOPIC INJECTION NEEDLE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
kaylin davis
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key11844036
MDR Text Key267674823
Report Number1037905-2021-00203
Device Sequence Number1
Product Code FBK
UDI-Device Identifier00827002216556
UDI-Public(01)00827002216556(17)211031(10)W4138564
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K851952
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberG21655
Device Catalogue NumberMHI-1-240
Device Lot NumberW4138564
Was Device Available for Evaluation? No
Date Manufacturer Received04/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-