• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SEASPINE ORTHOPEDICS CORPORATION SYNPLUG CEMENT RESTRICTOR; CEMENT OBTURATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SEASPINE ORTHOPEDICS CORPORATION SYNPLUG CEMENT RESTRICTOR; CEMENT OBTURATOR Back to Search Results
Model Number 804020
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Osteolysis (2377)
Event Date 03/29/2021
Event Type  malfunction  
Manufacturer Narrative
The product has been absorbed, and therefore no samples can be provided for testing or analysis.The patient will continue to be monitored by their healthcare provider for any indications which would necessitate intervention.Risk management files for the product have been updated to include the originally unidentified risk regarding the occurrence of osteolysis in some patients implanted with the biodegradable cement restrictors.Additionally, a risk benefit analysis was completed in may 2014 that concluded the overall medical benefits of the synplug products may not outweigh individual residual risk of osteolysis, primarily due to the lack of osteolysis reports for other cement restrictor types and options available to surgeon.Based on all available data at that time, no causal relationship between the use of the products and the reported instances of osteolysis could be demonstrated.To date, there are no conclusive findings from similar complaint investigations, or from technical or clinical information in the literature that proves or disproves a causal relationship between synplug & optiplug biodegradable cement restrictors or the materials they are manufactured with, and periprosthetic osteolysis (or fractures as a result).There is no data or findings that would suggest that only some subset of all the products manufactured might be affected.A risk/benefit analysis of two reported fractures, 57 reported osteolysis without the need for medical intervention and (b)(6) patients benefiting from a successful surgery it can be concluded that the benefit outweighs the risk.The finding of osteolysis surrounding the distal cement restrictor is unexpected, and undesirable; however, periprosthetic osteolysis in total hip arthroplasty is a well-known problem that is typically a multifactorial process and may be identified through routine radiographic follow-up.
 
Event Description
The patient underwent hip replacement surgery in (b)(6) 2006.In a 15-year postoperative follow up appointment on (b)(6) 2021, osteolysis was noted in the distal third of the femoral shaft.The patient does not report any pain or symptoms related to osteolysis.No further measures of intervention will be pursued at this time.The clinic continues to monitor the patient for any symptom necessitating treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNPLUG CEMENT RESTRICTOR
Type of Device
CEMENT OBTURATOR
Manufacturer (Section D)
SEASPINE ORTHOPEDICS CORPORATION
5770 armada dr.
carlsbad CA 92008
Manufacturer (Section G)
SEASPINE ORTHOPEDICS CORPORATION
5770 armada dr.
carlsbad CA 92008
Manufacturer Contact
macy nicol
5770 armada dr.
carlsbad, CA 92008
7602165137
MDR Report Key11852554
MDR Text Key252213633
Report Number3001503333-2021-00003
Device Sequence Number1
Product Code LZN
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K010840
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number804020
Device Catalogue Number804020
Was Device Available for Evaluation? No
Date Manufacturer Received04/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age84 YR
-
-