Model Number P7500A000131
Device Problems Intermittent Continuity (1121); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2021
Event Type  malfunction  
Manufacturer Narrative
The hillrom technical support representative performed troubleshooting over the phone with the account, asking the account to isolate and evaluate the pendant and siderails.It is necessary for the progressa¿ bed to have an effective maintenance program.We recommend that you do annual preventive maintenance examine the motors for the presence and tightness of the attachment hardware.Replace as necessary.Fully raise and lower the bed.Make sure there is no friction or abnormal noises and no audible overload indication can be heard during the movement.Make sure the bed not down led lights up on the control pendant and goes out when the bed is in low position.Replace the defective motor(s) in the event of a malfunction.Troubleshoot in the event of a doubt.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician at the account replaced the siderail caregiver controls to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hillrom technical support received a report from the customer stating the bed would move into chair position on it's own.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRESSA FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
emily mitchell
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key11855059
MDR Text Key251705263
Report Number1824206-2021-00252
Device Sequence Number1
Product Code FNL
UDI-Device Identifier00887761000100
UDI-Public010088776100010011201008
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP7500A000131
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-