• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problems Inability to Irrigate (1337); Expulsion (2933)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Visual examination of the returned product/provided pictures identified one of the batteries had leaked and crystalized onto the terminal and the through the wiring connected the terminal.Inspection of the interior of the hand piece found the same crystalized substance on the trigger electrodes and motor, and the motor was corroded and rusted.Functional testing found the device would not power on when connected to the original battery pack or the lab battery pack.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that water wasn't running from palsa vac during surgery.Investigation found battery leakage.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11855081
MDR Text Key251703537
Report Number0001526350-2021-00596
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00889024380042
UDI-Public(01)00889024380042(17)230430(10)64723001
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model NumberN/A
Device Catalogue Number00515047501
Device Lot Number64723001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2021
Was the Report Sent to FDA? No
Date Manufacturer Received05/13/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-