• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. WM CLARIS CPU; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. WM CLARIS CPU; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H700977
Device Problems Failure to Power Up (1476); Application Program Version or Upgrade Problem (2881)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation revealed that after the workmate claris (b)(4) software upgrade, the workmate claris dws screen turns black/blank and the user needs to reboot the system to regain functionality.
 
Event Description
Post network configuration claris 1.2 would not boot to application.The computer was taken off the domain to resolve the issue.There was no patient involved.The device was retained the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WM CLARIS CPU
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key11855202
MDR Text Key251698622
Report Number2184149-2021-00176
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05415067009777
UDI-Public05415067009777
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 11/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH700977
Device Catalogue NumberH700977
Device Lot Number7326257
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/19/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/07/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number2184149-05/05/21-001-R
Patient Sequence Number1
-
-