Model Number CD3365-40C |
Device Problems
Intermittent Capture (1080); Over-Sensing (1438)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Further information was requested but was not yet available.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that crosstalk was observed on the implantable cardiac defibrillator.
|
|
Event Description
|
New information received notes the patient was having 2-3 second pauses on telemetry due to pacemaker inhibition while in hospital following knee surgery.Programming changes to sensitivity were made and to prevent pauses and inhibition.The physician was to continue monitoring the patient.The patient did not state they were having any symptoms as a result of the observation.
|
|