Model Number 66800039
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/03/2021
Event Type  malfunction  
Event Description
It was reported that, the orange key could not turn in the socket of a versajet ii console.No case involved; therefore, no patient injuries or other complications were reported.No further information is available.
 
Manufacturer Narrative
The device intended for use in treatment was returned for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.Functional inspection was performed and showed the test pump cartridge could not be turned to the locked position due to corrosion inside u.I.Assembly which could establish a relationship between the device and reported event.The root cause is determined to be corrosion.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other failures related to the report event.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for adverse trends related to this product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET II CONSOLE
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11877600
MDR Text Key252426779
Report Number8043484-2021-01194
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00040565124759
UDI-Public00040565124759
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143115
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66800039
Device Catalogue Number66800039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received09/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-