DEPUY ORTHOPAEDICS INC US PIN REV DOME DRILL 35MM; DRILLS (BITS, SLEEVES, GUIDES) : DRILLS
|
Back to Search Results |
|
Model Number 2274-35-800 |
Device Problems
Break (1069); Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Sterile services informed drill bits were broken when found in the set.It broke at the point where the drill connects to the shaft.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: b3 and h6 (device code).
|
|
Event Description
|
Additional information was received, indicated that the part of the drill bit broke.According to sterile services description: at the point, where the drill connects to the shaft in one piece.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: the instrument associated with this report was not returned.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
|
Search Alerts/Recalls
|
|
|