|
Model Number OM-7000 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/1901 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference (b)(4).
|
|
Event Description
|
It was reported that the speedstitch device was not taking the needle.No case reported; therefore, there was no patient involvement.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).H10 h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual evaluation showed the device was not returned in any original packaging.The overall appearance of the device shows signs of wear.The suture cartridge has been incorrectly installed prior to the needle being installed.A functional evaluation showed the triggers will deploy the tissue jaw tube and lock into place.The suture cartridge is installed incorrectly and cannot be removed without damaging it or possibly the device.The needle cannot be installed due to the suture cartridge having been installed before the needle, which is in the wrong order.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed and the root cause was associated with unintended use of the device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|