OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problems No Device Output (1435); Output below Specifications (3004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2021
Event Type  malfunction  
Manufacturer Narrative
During troubleshoot, the technical support engineer confirmed the evis exera iii light source is compatible for use with the visera elite video system.The user facility had already connected the evis exera iii light source to the visera elite video system and it was working fine.The customer was provided lamp part number y1064sp to order replacement bulb via email.The investigation is ongoing; therefore, the root cause of the reported malfunction cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The technical support engineer was informed, during maintenance the visera elite xenon light source's main lamp reportedly was not working and the spare lamp ignited but was yellow in color and not providing adequate light.The customer does not have a compatible bulb replacement and inquired if they could use an evis exera iii light source with the visera elite video system.No patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11891820
MDR Text Key271046133
Report Number8010047-2021-06776
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/05/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-