MAUDE Adverse Event Report: KAVO DENTAL GMBH MASTERMATIC LUX M25 L; DENTAL HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KAVO DENTAL GMBH MASTERMATIC LUX M25 L; DENTAL HANDPIECE Back to Search Results
Model Number M25 L
Device Problems Mechanical Problem (1384); Overheating of Device (1437); Device Handling Problem (3265)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2021
Event Type  malfunction  
Manufacturer Narrative
The visual inspection prior to the repair showed that there was a hard, white substance sticking on the push button.It caused that the push button was not able to move free, hence it was blocked in the pushed position.This caused a permanent contact to the inner spinning parts which caused high friction and hence heat up.It is very likely that the white substance was a dental composite material which is used for dental restorations.The following test run showed that the running characteristics of the handpiece have been out of specification.The bearings have been running gritty and the power consumption was out of tolerance.This is a sign that the inner friction was higher due to worn out ball bearings.As the handpiece was only 5 month in use it is very likely that the blocked push button caused a stronger wear of the ball bearings.To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: warning: hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, untypical warming or when the cutter or grinder cannot be held.Do not use further and notify service.Caution: burning hazard from hot instrument head or hot instruments cover.If the instrument overheats, burns may arise in the oral area.Never contact soft tissue with the instrument head or instrument cover.The following guidelines must be observed to ensure save use of the electrically driven contra-angle handpieces: the service instructions for contra-angle handpieces must be precisely following when using kavo spray or quattrocare care systems.Before each use, the contra-angle handpiece must be checked for external damage.Before each use, perform a test run with the contra-angle handpiece, and watch for atypical heating and unusual noise and vibration.Immediately stop using contra-angle handpieces that act unusual.Never press the push button during operation.This also includes lifting the cheek or tongue! to ensure proper function, the medical device must be set up according to the reprocessing methods described in the kavo instructions for use, and the care products and care systems described therein must be used.Kavo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device.This service interval depends on the frequency of use and should be adjusted accordingly.
 
Event Description
The handpiece was sent in for repair and the accompanying document stated that the handpiece overheated during a treatment and burned the patient.During a call with the dental office it was described that the product got hot and the patient complained about the heat when the handpiece touched the lip.There was no visible injury after the appointment and therefore also no medical care necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERMATIC LUX M25 L
Type of Device
DENTAL HANDPIECE
Manufacturer (Section D)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM  88400
Manufacturer (Section G)
KAVO DENTAL GMBH
bismarckring 39
biberach / riss, 88400
GM   88400
Manufacturer Contact
klaus reisenauer
bismarckring 39
biberach / riss, 88400
GM   88400
MDR Report Key11892331
MDR Text Key252910446
Report Number3003637274-2021-00018
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K073478
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial
Report Date 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM25 L
Device Catalogue Number1.009.3630
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-