• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH UNKNOWN BIOMATERIAL - CEMENT; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH UNKNOWN BIOMATERIAL - CEMENT; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: 510k: this report is for an unknown cement/unknown lot.Part and lot number are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a synthes employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from the (b)(6) reports an event as follows: it was reported that there were 73 subjects in this report for a total of 84 patient.11 patients were not included in this final report because there were no follow up data.Treatment for confidence was 59 patient, perimeter 14 patient, and was not reported for the 11 patients.Confidence group had 39 females and 20 males with a mean age of 69.8 years.Perimeter group had 13 females and 1 male with a mean age of 75.7 years.Medical history reported: allergies, cancer/malignancy, cardio/circulatory, dermatological, endocrine/metabolic, gastrointestinal, hematological, musculoskeletal, neurological, psychiatric, respiratory, urogenital, diabetes, and osteoporosis.Study enrollment started on february 10, 2010, the last enrollment was february 21, 2011, and the last completed visit follow up was completed on february 24, 2011.The levels treated were between t7-l5.Twenty-five percent if subjects treated with confidence experience cement leakage.Thirty-six percent of leakage in perimeter cases.Cement leakage: confidence: 16 patients ¿ location 12 discal and 4 venous ¿ severity: 8 mild, 3 moderate, 5 none.Perimeter: 5 patients ¿ location 5 venous ¿ severity: 3 mild, 1 moderate, 1 none.One subject from both confidence and perimeter treated subjects experienced a new fracture at the one month follow up.There were three serious adverse events reported in the study however, none of these were device related.There was no device related adverse events reported during the procedure for either confidence or perimeter.One adverse event was reported for the confidence study as a discal cement leakage that was not device related or serious and was resolved.Subject 252 ¿ cement leakage (malfunction only no adverse event reported) subject 783 ¿ death ¿ unrelated to device this report is for an unknown cement.This is report 3 of 5 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN BIOMATERIAL - CEMENT
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
SENTIO (INNOVATIVE SURGICAL SOLUTIONS, LLC)
50461 west pontiac trail
wixom MI 48393
Manufacturer Contact
kara ditty-bovard
chemin-blanc 38
le locle 
SZ  
6103142063
MDR Report Key11893735
MDR Text Key266976673
Report Number1526439-2021-01052
Device Sequence Number1
Product Code NDN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,study
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/09/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2012
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-