Device evaluation by manufacturer: a field service engineering (fse) was at the customer's site to address reported event.Fse confirmed reported error by reviewing the error log but was unable to reproduce the error.Fse noticed a decrease in eki board liquid sensitivity and adjusted it which resolved the issue.Fse successfully validated instrument by performing quality control run without error and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no similar complaints identified during the search period.The aia-2000 operator's manual under appendix 4: error messages states the following: [2085] air detected during diluent suction by main arm cause: during diluent suction, it was determined that the tip failed to touch the liquid surface.If retry fails, the measurement result will be flagged (ds flag).Solution: contact tosoh service center or local representatives.The most probable cause of the reported event was due to liquid ad was on the low end of the range.
|