(b)(4), the sample was returned for evaluation.A visual exam was performed and the trim ring was found to be cracked.No other defects were observed.Functional testing was performed and the unit was connected to 120vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.No odor was noted.The unit was prepared for the functional bench test where a water reservoir bottle, an adult breathing circuit (880-36kit), a 382-10 conchasmart column, and dual temperature probes were connected to the unit.Air flow was supplied to the unit for a real time operational scenario.The unit successfully navigated all pre-operational self-tests again and was functioning in real time.The settings were set to full rainout, invasive, at 37 degrees c.The unit functioned without any interruption or functional anomalies for ~1.0 hour.No unexpected odors were noted.A device history record review was performed and no relevant findings were identified.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
|