Catalog Number 71440822
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/08/2021
Event Type  malfunction  
Event Description
It was reported that, when a tka surgery was finished the gii post ref a/p blk sz 7 had a missing pin.They checked the patient wound, the drapes and the floor but the missing part was not located.An xray was performed to determine if the part was in the patient but there was no sign on the xray.The procedure was completed, with a delay (less than 30 minutes) using the same device.No patient injuries and no other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Additional information: d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII POST REF A/P BLK SZ 7
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11902185
MDR Text Key253207878
Report Number1020279-2021-04760
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71440822
Device Lot Number10HM09660
Was Device Available for Evaluation? No
Date Manufacturer Received06/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-