SMITH & NEPHEW, INC. GII QUICK CONNECT HANDLE; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
Model Number 71440044 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/17/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during tka surgery, two gii quick connect handle will no longer engage and the hex screwdriver is stripped out.A smith and nephew back up device was available.No injury to patient or delay reported.
|
|
Manufacturer Narrative
|
H6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device has nicks and scratches on the device.The device shows signs of extensive use.A functional evaluation confirmed that the device was paired with mating part mechanism and does not function as intended.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|