SOFRADIM PRODUCTION SAS MESH SOFRADIM- PARIETEX COMPOSITE MESH; MESH, SURGICAL, POLYMERIC
|
Back to Search Results |
|
Model Number PCO1510 |
Device Problem
Degraded (1153)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during laparoscopic abdominal wall hernia procedure, when the mesh was fixed in the abdominal wall, it was found that there was a large area in the middle part that lacks anti-adhesion layer.The customer was concerned about the adhesion between organs in the later stage, so the mesh was replaced with a new one.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the mesh was found a little distorted: mesh dimensions found: 14 cm x 10 cm.The collagen film was found partially damaged (peeled off, torn and folded) on the overlap of the mesh and peeled off in a large half of the mesh.The collagen film was yellowed all over the remaining part.It was reported that the mesh was degraded or brittle.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|