Catalog Number 2C1008K
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/29/2021
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume infusor had a cracked luer connector.This was observed during patient infusion, when connected to a non-baxter luer activated valve.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The lot was manufactured from january 22, 2020 - january 24, 2020.The device was received for evaluation.A visual inspection was performed using the naked eye which observed a non baxter needle attached to the distal luer of the infusor and the tip of the luer has been damaged in which a piece from the luer was cracked and the piece d broken off.The reported condition was verified.The cause of the crack could not be determined; however, the most probable cause is user related or defective distal luer.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11916552
MDR Text Key258591047
Report Number1416980-2021-03269
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412022512
UDI-Public(01)00085412022512
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C1008K
Device Lot Number20A035
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2021
Date Manufacturer Received06/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment