NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number PS10.0-65D250
Device Problem Material Erosion (1214)
Patient Problem Pain (1994)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information has been provided, a supplemental report will be submitted.
 
Event Description
Information was received that the patient experienced pain at the junction of the nail.Upon removal, it was noted that there was rust at the junction and a huge callus formation.
 
Event Description
No additional information was provided.
 
Manufacturer Narrative
The nail was returned for visual and functional testing.Visual inspection revealed that the device had corrosion on the junction of the distraction rod.Functional testing was not applicable due to the reported failure mode.Device history review: a device history review (dhr) was performed on lot number: 9051314 and there were no deviations in the manufacturing process and the finished product met all acceptance criteria prior to shipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key11922080
MDR Text Key253787045
Report Number3006179046-2021-00288
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00887517951175
UDI-Public887517951175
Combination Product (y/n)N
PMA/PMN Number
K180503
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPS10.0-65D250
Device Lot Number9051314
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age14 YR
Patient Weight60
-
-