Model Number PS10.0-65D250 |
Device Problem
Material Erosion (1214)
|
Patient Problem
Pain (1994)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information has been provided, a supplemental report will be submitted.
|
|
Event Description
|
Information was received that the patient experienced pain at the junction of the nail.Upon removal, it was noted that there was rust at the junction and a huge callus formation.
|
|
Event Description
|
No additional information was provided.
|
|
Manufacturer Narrative
|
The nail was returned for visual and functional testing.Visual inspection revealed that the device had corrosion on the junction of the distraction rod.Functional testing was not applicable due to the reported failure mode.Device history review: a device history review (dhr) was performed on lot number: 9051314 and there were no deviations in the manufacturing process and the finished product met all acceptance criteria prior to shipment.
|
|