The field service engineer (fse) conducted a site visit and was able to confirm the error via the error log.The fse was able to reproduce the reported event by running test samples.Troubleshooting revealed sv171 valve intermittently not functioning correctly causing a drip to form/fluid to come out from number 2 probe tip.The fse replaced the sv171 to correct and prevent error.The fse confirmed liquid detection of probe 2 and ran washer prime many times without error.The fse ran four test samples without error.A daily check and qc was also performed successfully.The aia-900 analyzer is functioning as expected.No further action required by field service.A 13-month complaint history review and service history review for similar complaints were performed for serial number (b4) from 05apr2020 through aware date of 05may2021.There were 2 similar complaints identified during the searched period including this event.The aia-900 operator's manual under appendix 4: error messages states the following: error message: [2235] b/f probe 1 suction failure cause: the overflow sensor 1 s132 detected liquid after a water suction operation.A wu flag will be attached to the measurement result.Action: clean the wash probe 1.Check s132, the waste liquid solenoid valve sv170, the waste liquid tube, and the liquid pump lp172.The most probable cause of this reported event was due to failure of the valve.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|