The logbook provided was analyzed for the investigation.The described issue could be confirmed and no indications for a technical device malfunction were found.The case in question was started on (b)(6) 2021 at 4:38 pm in pressure mode.The following ventilation was accompanied by a permanent lack of fresh gas from the beginning, the device alerted accordingly.The flow measurement detected mv leaks of up to 3.1 l/min.As a result, there were strongly fluctuating pressure conditions with pronounced negative as well as positive pressure peaks.The device consequently alerted both aw.Pressure high as well as pressure negative.At 4:41 pm, a blockage of the piston occurred as a result of an unexpectedly rapid increase in pressure, whereupon the ventilator initiated an automatic shutdown as well as the autonomous switchover to man/spont.At the same time, ventilator failure was alerted.Ventilation continued in manual mode and was terminated at 4:54 pm with a switch to standby.Root cause for the reported symptom was an overpressure at the patient end of the circuit leading to a pressure peak.Most likely, the unstable pressure situation was the result of the patient breathing or coughing against the ventilator.To prevent from damages, the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.Dräger finally concludes that the device responded as specified upon the detected situation with an autonomous shutdown while changing mode to man/spont (safety mode) accompanied by an audible and visible "ventilator fail" alarm.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|