• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ZILVER PTX 35 DRUG-ELUTING STENT; NIU STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ZILVER PTX 35 DRUG-ELUTING STENT; NIU STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Catalog Number UNKNOWN
Device Problems Obstruction of Flow (2423); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Restenosis (4576)
Event Type  Injury  
Manufacturer Narrative
Exact rpn is unknown pma/510(k) #: p100022/s027 or p100022/s014.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event Description
The title is ¿the case treated with excimer laser for tandem lesions including repeated in-stent restenosis (re-restenosis) of des implanted in a sfa¿.The patient is a (b)(6) year-old female who has been visiting our hospital due to scleroderma.Intermittent claudication of both lower limbs and bilateral low abi had been observed since 2016, and zilver ptx was implanted in the right sfa-cto in august 2018.Intermittent claudication recurred from april 2019, and isr was observed in the right sfa, and poba was performed.In october 2020, the right lower extremity became cli, and angiography revealed re-restenosis in the stent, as well as new lesions in the proximal and progression of calcified nodules that had previously failed to dilate.We report the case because we performed treatment by using excimer laser mainly isr.
 
Event Description
Supplemental report is being submitted due to the completion of the investigation.
 
Manufacturer Narrative
Pma/510(k) #: p100022/s027 or p100022/s014.Device evaluation: the zisv6 device of unknown lot number involved in this complaint was implanted in the patient and was not available for evaluation.With the information provided, a document based investigation was conducted.Lab evaluation ¿ n/a.Document review: as the rpn and lot number of the complaint stents are unknown, a review of the relevant manufacturing records cannot be conducted.However, prior to distribution zisv6 devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.It should be noted that restenosis of the stented artery is listed as a known potential adverse event within the instructions for use (ifu0117-5).There is no evidence to suggest the user did not follow the ifu.Image review ¿ n/a.Root cause review a definitive root cause could not be determined from the available information.A possible root cause could be attributed to patient pre-existing/underlying conditions.From the literature article it is known that patient pre-existing conditions included scleroderma, intermittent claudication of both lower limbs and bilateral low ankle-brachial index.Restenosis of the stented artery is also listed as a known potential adverse event within the ifu and is a common adverse event of endovascular procedures that can be caused by injury to the vessel (e.G.During percutaneous transluminal angioplasty (pta) and/or stenting).Vessel injury provokes an inflammatory response that leads to or amplifies the restenosis process.It may be noted that the surface of the zilver ptx stent is coated with the drug paclitaxel to help prevent subsequent restenosis of the artery.Summary complaint is confirmed based on customer testimony.The patient outcome is unknown.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZILVER PTX 35 DRUG-ELUTING STENT
Type of Device
NIU STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
aisling hassett
o halloran road
national technology park
limerick 
MDR Report Key11937776
MDR Text Key267337384
Report Number3001845648-2021-00434
Device Sequence Number1
Product Code NIU
Combination Product (y/n)Y
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Hospital
Date Manufacturer Received05/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-