Device evaluation the zisv6 devices of unknown lot number involved in this complaint were implanted in the patients and were not available for evaluation.With the information provided, a document based investigation was conducted.Lab evaluation ¿ n/a document review as the rpn and lot number of the complaint stents are unknown, a review of the relevant manufacturing records cannot be conducted.However, prior to distribution zisv6 devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.It should be noted that ischemia requiring intervention ( bypass or amputation of toe, foot or leg) is listed as a known potential adverse event within the instructions for use (ifu0117-5).There is no evidence to suggest the user did not follow the ifu.Image review ¿ n/a root cause review a definitive root cause could not be determined from the available information.A possible root cause could be attributed to patient pre-existing/underlying conditions.From the literature article it is known that patient pre-existing condition was femoropopliteal disease.The patient outcome is unknown.Complaints of this nature will continue to be monitored for potential emerging trends.
|