Model Number EL5ML
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: unknown.Date of event is 2021.Event day and event month were not reported.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Upon visual analysis of the returned sample, it was determined that the el5ml device was returned unopened.Upon visual inspection of the tyvek, some cuts were observed on it, as if its cuts were performed with a cutter.Also, the blister was noted broken in one of the corners and the remainder was observed still adhered to tyvek.The reported complaint was confirmed.Due to the damages found on the blister, a possible cause for this condition is due to improper handling during transit or storage.It appears that the package hit a hard surface, and this caused the reported event.All ees product is 100% inspected prior to release.As part of our quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
 
Event Description
It was reported that the pvc package had obvious damage.Tyvek was not perforated.No additional information was provided.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
*  
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
*  
6107428552
MDR Report Key11953562
MDR Text Key255378805
Report Number3005075853-2021-03165
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberU94W0C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2021
Date Manufacturer Received05/18/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-