• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP HI LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP HI LITE ORANGE; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL Back to Search Results
Catalog Number 930815
Device Problems Defective Component (2292); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Pr (b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.
 
Event Description
Material no.: 930815.Batch no.: unknown.It was reported by the sales representative that the back end fell off causing the glass to break into the surgical field and onto patient.If the problem is corrected.Was new product shipped out to replace the product that was recalled.No date of incident was given.She should have it in their system broken glass and solution went into the surgical field and on to the patient.
 
Event Description
It was reported by the sales representative that the back end fell off causing the glass to break into the surgical field and onto patient.
 
Manufacturer Narrative
Your facility did provide not photos/samples to aid in our quality engineer¿s investigation.With the lack of samples provided, bd was unable to confirm the failure mode as the end cap was detached from the applicator body.A device history record could not be evaluated as the lot number is unknown.Although the complaint could not be confirmed, the root cause is attributed to the equipment station for the end cap placement unto the applicator body.Corrective actions were initiated which led to bd conducting a voluntary recall on certain lots of the chloraprep hi- lite orange 26 ml applicator.Bd has confirmed that some of the product had an applicator end cap that was improperly secured during the manufacturing process which resulted in broken glass dropping out of the applicator.H3 other text : see narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORAPREP ONE STEP HI LITE ORANGE
Type of Device
2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
Manufacturer (Section D)
CAREFUSION 213, LLC 0113
1550 northwestern dr
el paso TX 79912
MDR Report Key11956396
MDR Text Key267866234
Report Number3004932373-2021-00278
Device Sequence Number1
Product Code KXG
Combination Product (y/n)Y
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 08/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number930815
Device Lot NumberUNKNOWN
Date Manufacturer Received08/08/2021
Patient Sequence Number1
Patient Outcome(s) Other;
-
-