• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH Back to Search Results
Model Number M00545000
Device Problems Mechanical Problem (1384); Material Deformation (2976); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 05/10/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The suspect device was contaminated and is not available for return.If any further relevant information is received, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was used in the common bile duct area to diagnose a biliary stricture during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, the physician could not extend the brush outside of the catheter.After three attempts, the device was inspected and it was noted that the wire was kinked inside the catheter.They tried to push the brush out of the catheter when it was outside of the patient, but it did not come out.The device was not manipulated in any way before the procedure and it was not in a tortuous position when the problem occurred.There were no other issues noted with the device.The procedure was aborted and was not completed due to this event.There are no known patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key11956453
MDR Text Key254859068
Report Number3005099803-2021-02491
Device Sequence Number1
Product Code FDX
UDI-Device Identifier08714729268628
UDI-Public08714729268628
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/03/2023
Device Model NumberM00545000
Device Catalogue Number4500
Device Lot Number0026743178
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-